Direct Read Rotameters |
| FLD Series |
|
 |
|
|
|
|
$
95.00
FLDA3220ST
|
| |
Direct Read Rotameters for Air, Water, N2, H2, CO2, Ar, He, and O2
|
| |
Rib-Guided or Fluted Metering Tubes Facilitate Stable, Accurate Readings
|
| |
Magnifier Lens in Front Shields to Enhance Reading Resolution
|
| |
Interchangeability of Flow Tubes and Floats
|
| |
Ease of Installation and Exchange of Flow Tube
|
| |
Simple Means of Panel Mounting
|
View related products - Variable Area Flowmeters and Rotameters
|
| |
|
|
|
 |
Designed for direct read flow rates,
the FLD Series rotameter is a
precision instrument embodying the
inherent simplicity, versatility and
economy of the classical rotameter.
It is particularly suitable for
metering carrier gases in
chromatography, indicating and
controlling gases in manufacturing
processes, liquid and gas
measurement in laboratories, pilot
plants, flow and level indicating, etc.
Shipped completely assembled,
flow meters include standard
mounting fittings in a choice of
materials, side plates, thick
protective magnifying front shield
and back plate, built-in control
valve, and flow tubes. Panel
mounting style is convertible to
bench mounting through the use of
the optional acrylic tripod, FLTR-1.
The tripod has a built-in spirit
leveler and leveling screws.
Meters come with built-in 6 turn
needle valves, high precision
metering valves with “non-rising
stems” 16 turn, or with no valves.
The higher cost of the high
resolution valves is justified
whenever high sensitivity control
and resolution are desirable—
particularly in conjunction with
metering tubes of very low flow
rates. Generally, for gas metering it
is recommended that valves are
positioned at inlets (bottom) for
liquids valves may be positioned
either at inlets or outlets (top).
For vacuum services, valves
must be mounted at outlets. If
unspecified at the time of ordering,
meters will be shipped with valves
mounted at the inlets.
SPECIFICATIONS
Standard Accuracy: ±5% of full scale (FS)
Calibrated Accuracy: ±1% FS
Repeatability: ±0.25%
Useful Flow Range: 10:1 minimum
Maximum Operating Pressure: 200 psig/13.8 bars Maximum Operating Temperature: 121°C (250°F)
Construction Flow Tubes: Heavy walled borosilicate glass
Floats: Glass, sapphire, 316 stainless steel, carboloy and tantalum Choice of Mounting Fittings in Contact with Fluids: Aluminum, black anodized or 316 stainless steel
Side Panels: Aluminum, black anodized
Front Shield: Polycarbonate with longitudinal magnifier lens for enhanced reading resolution
Back Plate: 1⁄8" thick white acrylics
O-Rings and Packing: Buna O-rings in aluminum/brass model. FKM O-rings in stainless steel meters
Optional: Perfluoroelastomer and EPR
Connections: 1⁄8NPT female inlet and outlet connections
|
| |
 |
 |
|
 |
† All dollar amounts on this site are shown in US currency.
Note: “G” suffix denotes glass float, “ST” suffix denotes 316SS float, “C” suffix denotes carboloy float and “S” denotes sapphire float. For units without valve, add suffix “-NV” to model number and subtract $10 for aluminum; subtract $13 for for 316SS. 1. Minimum flow rate is 10% of the maximum rate shown. 2. For 16 turn, high resolution non-rising stem valve for 150 mm rotameters, add suffix “-HRV” to model number and add $50 for aluminum units and $100 for stainless steel units. (OPTIONS SHOWN IN THIS "NOTE" ABOVE ARE PRICED IN US$)
Ordering Example: (1) FLDN3501ST 6 turn valve stainless steel unit, with 316SS float, range of 1.6 scfm nitrogen = $247.00
|
|
 |
|
 |
|  |
|
Keywords.:rotameter, variable area flowmeter, flow meter, air flowmeter, water flowmeter, nitrogen flowmeter, argon flowmeter, tube and float, direct read flowmeter, direct read rotameter, flow measurement
|