The OM-CP-CRYO-TEMP ultra low
temperature data logger records
temperatures as low as -86°C
(-122ºF). The OM-CP-CRYO-TEMP
can be used in a broad line of
applications such as monitoring
blood plasma, vaccines,
pharmaceutical, frozen foods
and shipping containers.This
stand-alone device does not
require any additional probes.
The enclosure is designed with
a handle for easy attachment and
is IP64 splash resistant.
It utilizes a USB docking station
(sold separately), to communicate
with a personal computer or laptop.
The OM-CP-CRYO-TEMP provides
date and time stamped temperature
readings. Our easy-to-use Windows
software does the rest.
The user can view the data in
graphical or tabular form. High and
low alarm limits and high and low
warning limits can be programmed
through the software.
The user can set the alarm limits in
the range required for the goods to
be maintained. The warning limits
can be set to alert the user when
the temperature is nearing a high or
low alarm limit, so that proper action
can be taken to ensure the integrity
of the goods.
The OM-CP-CRYO-TEMP is
equipped with three LED’s. The
green LED will blink every 15
seconds while the recorder is
logging data. The yellow LED will
blink every 3 seconds when the
temperature has breached the
warning limits. The red LED will
blink every 3 seconds when the
temperature alarm limits have
been exceeded.
The OM-CP-CRYO-TEMP is the
ideal temperature monitoring
solution for any application involving
ultra low temperature recording.
Specifications Temperature Sensor: NTC Thermistor Temperature Measurement
Range: -86 to 35°C (-122 to 95°F)
recommended, may be used up to 55°C (131°F) Temperature Resolution: 0.1°C Calibrated Accuracy: ±1.0°C Specified Accuracy Range:
-80 to 0°C (-122 to 32°F) Readings: 32,767 Recording Interval: 5 seconds to 30 minutes,
selectable in software Start Modes: Immediate start, delay start or manual start
with magnetic switch Programmable Start Delay: Up to 30 days from PC launch
or manual start Temperature Alarm: Software programmable high and low alarm
limits, high and low warning limits; each setpoint may be individually
enabled or disabled Alarm Delay: Software programmable time delay for
warning/alarm limits; logger will not indicate out of range status until the
temperature has been out of range for a set period of time Status Indicators: Green: Blinks to indicate safe conditions during logging Yellow: Blinks to indicate warn limits have been exceeded Red: Blinks to indicate alarm criteria (limits/delay) have
been exceeded Password Protection: An optional password may be
programmed into the device to restrict access to configuration
options; data may be read out without the password Calibration: Digital calibration is available
to the user through software Calibration Date: Automatically recorded
within device Power: 3.6 V lithium battery included), factory replaceable only Battery Life: 2 Years: Typical @ 25°C, 1 min
reading intervals 1 Year: Typical @ -80°C, 1 min
reading intervals Time Accuracy: ±2 minutes/month max at 20 to 30°C;
±5 minutes/month max at -80°C Data Format: Date and time stamped °C, °F, K, °R Computer Interface: USB docking station to PC; 38,400 baud Software: XP SP3/Vista and 7 (32 and 64-bit) Operating Environment:
-86 to 35°C (-122 to 95°F)
may be used up to 55°C (131°F); 0 to 100% RH Dimensions: 83 H x 54 W x 11 mm D
(3.3 x 2.1 x 0.4") Ingress Protection/IP Rating: IP64 (splash resistant,
not submersible) Enclosure: ABS plastic, white,
polycarbonate labeling Weight: 40 g (1.4 oz)
† All dollar amounts on this site are shown in US currency.
Note: Comes complete with 3.6V lithium battery. Operator’s manual and interface cable are included with the OM-CP-IFC200 software/cable package (required for data logger operation, sold separately). Ordering Example: (1) OM-CP-CRYO-TEMP-CERT Ultra-low temperature data logger with NIST calibration certificate, $259.00