Omega.com - You're One-Stop Source for Process, Measurement and Control! (888) TCOMEGA (in US/Canada) or (203) 359- 1660 (for international)
Navigation Image Map  
search             

Toroidal Conductivity System

CDTX-45T  
New!
CDTX-45T:Toroidal Conductivity System
Click here for larger image.
   
Loop-Powered, AC, or Battery Versions
Dual Alarm Relays/Analog Outputs
Large, 2-Line Display

View related products - Conductivity Instrumentation
Scroll below to order View spec PDF

The CDTX-45T toroidal conductivity system is designed for online monitoring of chemically aggressive process solutions, water applications that coat or foul traditional sensors, and plating bath operations in which high current densities are present. The sensor consists of 2 metallic ribbon coils that are fixed in place by the sensor jacket material. The drive coil is used to induce a current in the process solution. The second sensing coil is used to measure the current in the process solution; the magnitude of this current is proportional to the conductivity of the process solution. Toroidal sensor is available in Noryl to ensure that sensor-wetted materials are chemically resistant to process chemicals. The sensor material is non-conductive, thereby isolating the sensor from electrical noise and ground loops that can influence the integrity of the measurement. This material also makes the sensor resistant to electrode coating, since most substances do not adhere to the sensor material. Sensor can be submersion mounted for easy installation in open tanks or pipe mounted in the optional 2" tee fitting, which is keyed to the sensor for proper alignment. Tee fitting material is polypropylene. For larger pipe diameters, the sensor can be insertion mounted through a 11/2 NPT ball valve for easy retraction. The CDTX-45TC is the concentration monitor. The monitor comes with concentration/temperature tables for sodium hydroxide, potassium hydroxide, hydrochloric acid, nitric acid, and sulfuric acid. A configurable table allows the user to enter data points for a custom concentration curve.

MONITOR PERFORMANCE SPECIFICATIONS
Measuring Range:
    Main Input: 0 µS to 2000 mS % concentration, mg/L total dissolved solids; loop current 4 to 20 mA
    Sensor Temperature: -10 to 210ºC (14 to 410ºF)
    Main Parameter Ranges: 0 to 2000 µS, 0 to 2 mS, 0 to 20 mS, 0 to 200 mS,
    0 to 2000 mS, 0 to 2 S
Repeatability: 0.3% of span or 0.1 µS, whichever is greater
Sensitivity: 0.05% of span or 0.1 µS, whichever is greater
Non-Linearity: 0.3% of span or 0.1 µS, whichever is greater
Stability: 0.1% of span per 24 hr
Temperature Drift: Span or zero, 0.03% of span/ºC
Warm-Up Time: 7 s to rated performance
Response Time: 6 s to 90% of step input at lowest setting Max Sensor-to-Analyzer Distance: 61 m (200')

SENSOR SPECIFICATIONS
Sensor: Toroidal electrode
Wetted Materials: Noryl
Sensor Cable: 6.1 m (20') standard, maximum length 61 m (200')
Response Time: 2 s to 90% FS
Temperature Range: 0 to 105°C (32 to 221°F)
Pressure Limit: 150 psig
Process Connection: 3/4 MNPT rear thread, optional 2" keyed tee fitting
Temperature Element: Pt1000 RTD
Minimum Conductivity: 500 µS

INSTRUMENT SPECIFICATIONS
Enclosure:
    Materials: NEMA 4X (IP66) polycarbonate, stainless steel hardware,
    weatherproof and corrosion resistant
    Dimensions: 112 H x 112 W x 89 mm D (4.4 x 4.4 x 3.5")
Mounting Options: Wall, panel (115 x 115 mm) or pipe/handrail
Conduit Openings: 3 M16 openings; cord grips and conduit adaptors included
Weight:
    Loop-Powered Transmitter: 0.45 kg (1 lb)
    Other Configurations: 0.68 kg (1.5 lb)
Display: Large, high-contrast, super-twist (STN) LCD 4-digit main display with sign, 19.1 mm (0.75") 7-segment characters; 12-digit alphanumeric second-line display, 7.6 mm (0.3") 5 x 7 dot matrix characters
Keypad: 4-key membrane type with tactile feedback, polycarbonate with UV coating, integral EMI/static shield and conductively coated window
Ambient Temperature:
    Operating: -20 to 60ºC (-4 to 140ºF)
    Storage: -30 to 70ºC (-22 to 158ºF)
Ambient Humidity: 0 to 95% RH, non-condensing
     Designed for hazardous and non-hazardous areas
EMI/RFI Influence: Designed to EN 61326-1
Output Isolation: 600 V galvanic isolation
Filter: Adjustable, 0 to 9.9 minutes additional damping to 90% step input
Temperature Input: Selectable Pt1000 or Pt100 RTD with automatic compensation
Please download specification page for complete list of available models.


To Order (Specify Model Number)    
 Part Number   Availability   Price    Description    Qty.
 Conductivity Analyzer
 CDTX-45T1 in-stock    $675.00     Loop powered 16 to 35 Vdc, 2-wire (4 to 20 mA output)  
 CDTX-45T2 in-stock    $925.00     120 Vac, 2 relays, two 4 to 20 mA outputs  
 CDTX-45T4 in-stock    $925.00     Battery operated (9 V), two 0 to 2.5 Vdc outputs  
 Sensor with 20' Cable
 CDE-45T2 in-stock    $405.00     Noryl electrode  
 Accessories
 CDTX-45T-JB in-stock    $95.00     NEMA 4X (IP66) junction box  
 CDTX-45T-C(*)     Consult sales   Sensor-interconnect cable  
 CDTX-45T-PT in-stock    $95.00     2" keyed slip-fit flow cell for use with polypro sensors  
 CDTX-45T-SMK in-stock    $160.00     Submersible mounting kit for PEEKTM sensor  
 CDTX-45T-PMK in-stock    $45.00     Panel mount kit for analyzer  
All dollar amounts on this site are shown in US currency.

 Note: Comes with complete operator's manual.
For extended cable lengths in foot increments up to 61 m (200'), add "-*" to model number and $3/ft to price.
(OPTIONS SHOWN IN THIS "NOTE" ABOVE ARE PRICED IN US$)
 Ordering Example: (1) CDTX-45T2 conductivity analyzer = $925

   
Keywords:Toroidal Conductivity System

   Back to top
© Copyright 2003-2008  |  OMEGA ENGINEERING, Inc.
OMEGA
Privacy Statement  |  Terms and Conditions
   MXV MMV | 72
VerisignAmex, Visa, Mastercard
Skip to navigation


Home Quick Order Products Literature My Omega Omega Info Contact Us Cart Help Global