The CL130 thermocouple calibrator has been designed
using the latest technology to provide reliable accuracy
throughout its entire operating temperature range. The
compact design makes the unit easy to carry and the
intuitive user interface makes it easy to use. Temperature
measurements/simulations are displayed on a large, easyto-
read liquid crystal display (LCD). This unit has two
backlights: a green backlight which is user activated to
improve readability in low light conditions, and an automatic
red backlight which indicates an over-range condition.
Twelve keys provide access to the functional modes and
menus, including measure, simulate, setup, run, record,
edit, and backlight activation.
SPECIFICATIONS mV Range, Accuracy: -15 to 80 mV, 0.005% Rdg ±6 mV mV Temp. Effect: ±2 µV/°C (Ref. Temp = 23°C) Cold Junction Effect: ±0.05°C/°C Cold Junction Accuracy: ±0.1°C at 25°C Resolution: Thermocouple Types E, J, K, N, T: 0.1°C Thermocouple Types B, R, S: 1°C mV: 0.001 mV (1 µV) Environmental: Storage: -40 to 85°C (-40 to 185°F) Operating: -10 to 50°C (14 to 122°F) Humidity: 10 to 90% RH Maximum Input Voltage: 30 Vdc Power Requirement Batteries: 3 “AA” batteries (included), field replaceable Dimensions: Enclosure: 165 H x 81 W x 28 mm D (6.5 x 3.2 x 1.1") Polycarbonate, Permanently Static Dissipative, ESD Protection Enclosure with Boot: 173 H x 89 W x 33 mm D (6.8 x 3.5 x 1.3")
† All dollar amounts on this site are shown in US currency.
Note: Comes complete with operator’s manual, shock resistant holster, 3 “AA” alkaline batteries. Ordering Example: (1) CL130-NIST Thermocouple/mV calibrator with NIST certificate, $1,022.00