Integral Signal Conditioners for Frequency, Voltage, or Current Output |
| FLSC-28, FLSC-34, FLSC-35 and FLSC-51 |
|
 |
|
|
|
|
|
|
|
| |
Up to 0.05% Accuracy
|
| |
Choose from Six Output Types
|
| |
CMOS/TTL Compatible
|
| |
Adjustable Low Level Signal Filter for Noise Rejection
|
| |
Explosion-Resistant Heads
|
| |
For FTB-100 and FTB-200 Turbine Flowmeters
|
| |
Integral Signal Conditioners Available for 4 to 20 mA, 0-5 V, and Factored Pulse Outputs
|
View related products - Turbine and Paddle Wheel Flowmeters
|
| |
|
|
|
 |
| Signal Conditioners | Compatible Meters Model No. |
FLSC-34, FLSC-28, FLSC-35B, FLSC-51 | DPF50, DPF60, DPF300, DPi8 |
| FLSC-51 | DPF70, DPF701 |
| See related links below. |
OMEGA® high accuracy integral signal conditioners are designed for direct mounting onto OMEGA
FTB-100 and 200 Series turbine meters. The input circuitry of the signal conditioners has been
designed to receive and condition low level turbine meter signals while rejecting unwanted noise and
spurious signals. A signal threshold control is provided which allows the user to set the input
sensitivity above the ambient noise level, thereby eliminating any false signal on the output. Signal
conditioners are available that can provide 4 to 20 mA, 0 to 5V, and scaled frequency pulse outputs
that run on unregulated power supplies. Other units which run off 115 Vac provide both an unscaled pulse
output and a 0 to 5V or 4 to 20 mA output. A zero and a span potentiometer allow for simple field
adjustment of the analog outputs from the signal conditioners.
SPECIFICATIONS
Accuracy: See "To Order" box
Temperature Range: 0 to 70°C (32 to 158°F)
Pulse Output: TTL/CMOS; open collector for FLSC-51B
Input Frequency: 10-1000 Hz, 28 mV P-P; 1000-2500 Hz, 70 mV P-P.
Can be used with any OMEGA Turbine Meter
Maximum Input: 120V RMS
Electrical Connections: Screw terminals
Maximum Lead Length: 305 m (1000') of 24 gage copper wire for pulse or mA;
30 m (100') for 0 to 5V output.
Weight: 2.3 kg (5 lb)
| Model No. |
Output |
Power |
Accuracy |
| FLSC-34 |
4 to 20 mA and unscaled pulse |
115 Vac, 50/60 Hz |
±0.1% FS |
| FLSC-28 |
0 to 5 Vdc |
10 to 40 Vdc @ 10 mA |
±0.5% FS |
| FLSC-35B |
0 to 5 Vdc and unscaled pulse |
115 Vac, 50/60 Hz |
±0.1% FS |
| FLSC-51 |
scaled squarewave pulse (2 ms duration) |
8 to 35 Vdc @ 10 mA |
±1% FS |
| FLSC-51B |
scaled squarewave pulse (50 ms duration) |
8 to 35 Vdc @ 10 mA |
±1% FS |
|
| |
 |
|
|
MOST POPULAR MODELS HIGHLIGHTED! |
|
 |
|
 |
† All dollar amounts on this site are shown in US currency.
Note: Signal conditioners are installed at no charge when purchased with turbine meter. Ordering example with turbine meter :SYS/FTB-101/FLSC-28 Add cost of turbine and signal conditioner together for system cost. NOTE: The use of explosion-proof heads does not imply that complete assembly is either explosion-proof or intrinsically safe. It is the user’s responsibility to determine the suitability of materials for a given installation.
Ordering Example: (1) FLSC-28 0-5 V output signal conditioner powered by 10-40 Vdc, $603.00
|
|
 |
|
 |
|  |
 |
 |
 |
 |
 |
 |
 |
 |
 |
Related Links -
Related Products
|
 |
 |
 |
 |
 |
 |
|
Related Products
|
 |
FTB-100:
|
|
Turbine Flowmeters with Economical Ball Bearing Design
|
 |
 |
|
 |
FTB-200:
|
|
Economical Ball Bearing Turbine Flowmeters With 37° Flare Fittings
|
 |
 |
|
|
|
 |
DPF50:
|
|
3 1/2 Digit Mini-size Process Meter
|
 |
 |
|
 |
DPF60 Series:
|
|
DPF60 Series Ratemeters, Totalizers and Batch Controllers for Voltage and Current Inputs
|
 |
 |
|
|
|
 |
 |
|
|