This flowmeter is precise and
sophisticated. It must be accurately
calibrated for application conditions.
If in doubt, consult OMEGA’s Flow
Department.
OMEGA® FMA-8500 Series high
pressure mass flowmeters feature
a superior ±0.25% repeatability for
accurate measurement and control
of gas flow. The flow sensor produces
a 0 to 5 Vdc output signal linear to the
mass flow rate for use in recording,
indicating, or control purposes.
SPECIFICATIONS
Accuracy: ±1% full scale at 16 to 27°C (60 to 80°F) and 10 to 40 psig; ±2% full scale at 4 to 38°C (40 to 100°F) and 5 to 150 psig Repeatability: ±0.25%
Time Constant: Three seconds to reach 98% of final value
Max Pressure:Up to 4500 psi at 66°C (150°F)
Pressure Drop: 10" water, max
Ambient Temperature Range: 4 to 66°C (40 to 150°F)
Output: 0 to 5 Vdc into 1000 Ω min load; 3 mV rms max ripple
Power: ±15 Vdc; 0.6 W max power consumption
Fittings: ¼" 316 SS compression fitting O-Rings/Gaskets: PFA Transducer Assembly: 316 SS
Leak Rate: 10-9 atm/cc-Helium
Pressure Coefficient: 0.03% shift within ±200 psig of the calibrated pressure
Range
Model No.
Range
FMA-8500-(*)
0 to 10 SCCM
FMA-8501-(*)
0 to 20 SCCM
FMA-8502-(*)
0 to 50 SCCM
FMA-8503-(*)
0 to 100 SCCM
FMA-8504-(*)
0 to 200 SCCM
FMA-8505-(*)
0 to 500 SCCM
FMA-8506-(*)
0 to 1000 SCCM
FMA-8507-(*)
0 to 2000 SCCM
FMA-8508-(*)
0 to 5000 SCCM
FMA-8509-(*)
0 to 10 SLM
FMA-8510-(*)
0 to 20 SLM
(*) You must specify gas and inlet/outlet pressure when ordering.
1) Select the core part number from the range code table above.
2) Add the gas and inlet/output pressure information after
the core p/n, eg.: FMA-8501-Nitrogen,20/0 psig (This is a FMA-8501 0 to 20 SCCM, flowmeter for nitrogen, with an inlet pressure of 20 psi, and outlet pressure of 0 psi)
3) Enter the Constructed part number in the "Order Specific Models Here" table below!
4) You will then add the p/n to your cart and check out with a non-std p/n.
5)We will confirm by email and/or telecon price and availability after the order is placed.
† All dollar amounts on this site are shown in US currency.
Note: Comes complete with mating connector and operator's manual.
Calibrations done at ambient temperature only 20°C(70°F).
* Please specify gas and inlet/outlet pressure. When ordering, replace the "*" in the suffix with the gas and inlet/outlet pressure information.
For calibrations between 201 and 900 psig; add $600 to price.
For calibrations between 901 and 4500 psig; add $675 to price.
For NIST certificate, add suffix "-NIST" to the model number; add $625.00 to price. NIST certificate does not include points.
All models are custom, consult specification page for complete description and list of available models.