The FMA-A2100/2200/2300/2400 Series Electronic Mass Flowmeters/ Controllers provide high performance,
versatility and state-of-the-art design in one compact package. The FMA-A2100/ 2200/2300/2400 uses
capillary-type thermal technology to directly measure mass flow of gases. No temperature, pressure,
or square root corrections are required. The FMA-A2300/2400 Series comes with an LCD display, and all
models come with linear 0-5 and 4-20 mA output.
The FMA-A2100/2200/2300/2400 measures the mass flowrate of gases in 24 ranges from 0-10 SCCM to 0-100
SLM. For the complete listing, refer to the range table below. The FMA-A2100/2200/ 2300/2400 Series
is compatible with most noncorrosive gases. The user is advised to check the wetted materials against
gas compatibility.
OMEGA’s Mass Flow Controllers use an internal electromagnetic proportional valve to control the mass
flowrate. A command signal can be supplied either by the internal setpoint pot or by a 0-5 Vdc external
source.
SPECIFICATIONS Accuracy: ±1% of full scale including linearity Repeatability: ±0.15% full scale or better Turndown Ratio: 100:1 Response Time: 5 seconds Gas Ambient Temperature: 0 to 50°C (32 to 122°F) Temperature Coefficient: 0.05% full scale per 1°C or better Pressure Coefficient: 0.01% full scale per psi or better Minimum Pressure Requirements: Meters: 1 inch of H2O Controllers ≤1 SLM: 1 psi Controllers >20 SLM: 20 psi Differential Pressure for Mass Flow Controllers: <5 LPM: 5 to 50 psi standard >5 LPM: 20 to 40 psi standard Maximum Gas Pressure: Polyacetal: 250 psig 316 SS: 500 psig Leak Integrity: 1 x10-6 cc/min of He or better Wetted Parts: Standard: Polyacetal, FKM O-rings Optional: 316 SS, FKM O-rings Connections: 1/4 FNPT standard; 1/4" compression fitting for all stainless steel models Input/Output Connection: 9-pin sub-D connector Control Signal: 0 to 5 Vdc Control Valve: Electromagnetic normally closed Output signal: Linear 0 to 5 Vdc, 1000 Ω min load, 4 to 20 mA, 500 Ω loop resistance Input Power: Meters: 24 Vdc @ 150 mA, 115 Vac, 220 Vac with AC adaptor Controller: 24 Vdc @ 300 mA, 115 Vac, 220 Vac with AC adaptor Dimensions: See drawing Weight: 1.25 kg (2.75 lb) with power supply
Range Table
Code
SCCM
Code
SLM
Code
SLM
00
0-10
06
0-1
14
0-35
01
0-20
07
0-2
15
0-40
02
0-50
08
0-5
16
0-45
03
0-100
09
0-10
17
0-50
04
0-200
10
0-15
18
0-55*
05
0-500
11
0-20
19
0-60*
12
0-25
20
0-65*
13
0-30
21
0-70*
22
0-75*
23
0-100*
*Only for FMA-A2100/2300 Series
*Insert range code from table at top of page to complete model number.
**Specify gas and inlet/outlet Pressure
FMA-A2200-SS-(**) and FMA-A2400-SS-(**) controllers have a stainless steel body (with 1/4" compression fittings)standard.
† All dollar amounts on this site are shown in US currency.
Note: Comes complete with 9-pin sub-D connector, power supply, NIST certificate (no data points), and operator's manual.
For a 5-point NIST certificate, add suffix "-NIST5" to the part number; no additional cost.
For a 10-point NIST certificate, add suffix "-NIST10" to the part number; add $125 to price.
Calibrations done at ambient (20°C or 70°F) temperature only.
CONTROLLERS: All controller flow ranges specified are for Nitrogen or air at 20; psig when used for other gases, a correlation factor is needed to determine the flow rate.
For non-standard calibrations, replace controller suffix "-(**)" with the desired gas and pressure.
METERS: Meter flow rates are specified at Nitrogen 14.7 psia.
** For optional 316 SS body meters, add "-SS" to model number; add $440 to price.
(OPTIONS SHOWN IN THIS "NOTE" ABOVE ARE PRICED IN US$)