Analog Gaging Probes |
| GP911 Series |
|
 |
|
|
|
 |
The GP911 series analog gaging probes utilize proven LVDT technology to provide excellent linearity and
repeatability in a compact, reliable unit. All analog probe heads carry a NEMA 12 (IP65) environmental rating
which makes them suitable for use in harsh industrial environments.
In an analog gage probe, two small transformers share a common magnetic core. As the core moves,
the output of one transformer increases and the other decreases. The out-of-balance current provides the
measure of the core’s position. LVDT’s operate on an ac energizing voltage and require special signal
conditioning.
SPECIFICATIONS
Excitation Voltage: 1 to 10 volts rms
Excitation Frequency: 2 to 20 kHz
Linearity: 0.5% rdg or 0.1% FS, whichever is greater
Repeatability: 0.15 microns
Calibration Voltage: 3 Vrms
Calibration Frequency: 5 kHz
Calibration Load: 10 kΩ
Storage Temp.: -40 to 100°C (-40 to 212°F)
Operating Temp.: -10 to 80°C (14 to 176°F)
Temp. Coefficient: 0.01% FS/°C (0.03% for 0.5 mm model)
|
| |
 |
|
 |
|
 |
† All dollar amounts on this site are shown in US currency.