The OM-MLT Series is a
comprehensive family of portable
data loggers that uses the latest
electronic technology and software.
These data loggers are compact,
self-powered devices that, using
a variety of internal and external
sensor types, will measure and
record almost every possible critical
condition. The OM-MLT Series also
includes a high-temperature model
and an autoclave model.
The OM-MLT-AC autoclave
data logger is a special product
that can withstand high-temperature
and pressure environments to
150°C (302°F)/5 bar for 30 minutes,
150°C (302°F) for 60 minutes in
dry heat and indefinite operation
at 125°C (257°F). The primary
intended use for the data logger is
the qualification of autoclaves used
for sterilization, however, it is also
suitable for measuring any high-temperature
applications not
exceeding the specifications.
It is supplied with a stub probe
that registers the environment
temperature from 25 to 150°C
(77 to 302°F), and an internal
temperature sensor for monitoring
of data logger internal temperature
conditions and prohibiting
reprogramming of data logger until
the internal temperature has
returned to 40°C (104°F).
Data logger setup and data
transfer to the PC is accomplished
using the OM-MLT Series
Windows software. Data loggers
can be programmed to start based
on date/time, or they can be
manually started by using the
starting magnet. The logging
session can end based on date/time
or total number of readings or it can
be set to wrap around. To download
your data, simply place the data
logger in the PC interface cradle
and connect to the data logger
using the OM-MLT Series Windows
software. The software performs
statistical analysis including
minimum, maximum and average
of all readings. The software also
features user defined alarm set
points and graph labels, a multiple
graphing facility to graph data from
multiple data loggers on a single
graph and also ability to view
current conditions through the PC
interface in real time. Stored data
files can be exported to Microsoft
Excel.
Specifications
COMMON SPECIFICATIONS
Sampling Rate: User selectable
5 to 59 sec or 1 to 255
min (OM-MLTHB16 is 1 to 255
min only) Memory: All models store
16,000 readings except
OM-MLT-AC, which stores
2000 readings Power: 3.6 V lithium battery
(included) Battery Life: 1 to 2 year battery life
typical depending on use Software: WIN95/98/NT/2000/XP
OM-MLTD16
TEMPERATURE DATA LOGGER
WITH ONE INTERNAL SENSOR
Temperature Range: -40 to 70°C
(-40 to 158°F) Accuracy: ±0.3°C (±0.5°F) Resolution: 0.5°C (0.9°F) Visual Indicators: Green LED
indicates active status, Red LED
indicates alarm Storage Temperature:
-40 to 85°C (-40 to 185°F) Enclosure: IP67 Weight: 70 g (2.5 oz) Dimensions:
75 D x 18 mm H
(2.95 x 0.7")
OM-MLTXD16
TEMPERATURE DATA LOGGER
WITH ONE INTERNAL AND ONE
EXTERNAL SENSOR
Temperature Range:
-40 to 70°C (-40 to 158°F) Accuracy: ±0.3°C (±0.5°F) Resolution: 0.5°C (0.9°F) Enclosure: IP51 Temperature Probe: 4.5 D x 80
mm L (0.18 x 3.1") Stainless steel
penetration style probe with
1 m (3') cable Visual Indicators: Green LED
indicates active status, Red LED
indicates alarm Storage Temperature:
-40 to 85°C (-40 to 185°F) Weight: 70 g (2.5 oz) Dimensions: 75 D x 18 mm H
(2.95 x 0.7 ")
OM-MLTXXD16
TEMPERATURE DATA LOGGER
WITH ONE INTERNAL AND TWO
EXTERNAL SENSORS
Temperature Range:
-40 to 70°C (-40 to 158°F) Accuracy: ±0.3°C (±0.5°F) Resolution:0.5°C (0.9°F) Enclosure: IP51 Temperature Probe: 4.5 D x 80
mm L (0.18 x 3.1") Stainless steel
penetration style probe with
1 m (3') cable Visual Indicators: Green LED
indicates active status, red LED
indicates alarm Storage Temperature:
-40 to 85°C (-40 to 185°F) Weight: 70 g (2.5 oz) Dimensions: 75 D x 18 mm H
(2.95 x 0.7")
OM-MLTHB16
TEMPERATURE/HUMIDITY
DATA LOGGER
Temperature Range: -20 to 60°C
(-4 to 140°F) Temperature Accuracy:
±0.3°C (±0.5°F) Temperature Resolution:
0.3°C (0.5°F) Humidity Range: 0 to 100% RH Humidity Accuracy: ±3% RH Humidity Resolution: 0.5% RH Visual Indicators: Green LED
indicates active status, Red LED
indicates alarm Enclosure: IP51 Storage Temperature: -40 to 85°C
(-40 to 185°F) Weight: 100 g (3.5 oz) Dimensions:
75 D x 50 mm H
(2.95 x 2.0")
OM-MLT-HT
HIGH TEMPERATURE
DATA LOGGER
Temperature Range:
0 to 125°C (32 to 257°F) Accuracy: ±0.3°C (±0.5°F) Resolution: 0.5°C (0.9°F) Enclosure: 316 stainless steel
with IP68 rating Storage Temperature:
-40 to 125°C (-40 to 257°F) Weight: 250 g (8.8 oz) Dimensions: Data logger:
49 D x 28 mm H (1.93 x 1.1") Probe: 4.5 D x 50 mm L
(0.18 x 2.0")
OM-MLT-AC
AUTOCLAVE DATA LOGGER
Temperature Range:
-25 to 150°C (-13 to 302°F) Endurance Ratings: Rated for
autoclave at 150°C (302°F)/5 Bar
for 30 minutes, typical oven dry heat
at 150°C (302°F) for 60 minutes,
and oven dry heat at 125°C (257°F)
for indefinite amount of time
(depending on battery life) provided
that temperature never exceeds
125°C (257°F) Sensors: External sensor is used
for sampling the environment under
test; internal sensor has a
measuring range of 1 to 150°C
(34 to 302°F) and is used for
diagnostic purposes only Accuracy: ±0.5°C (±0.9°F) Resolution: 0.5°C (0.9°F) Enclosure: 316 stainless steel IP68
tested to 5 bar
Storage Temperature: 0 to 125°C
(32 to 257°F) Weight: 345 g (12 oz) Dimensions: Data logger:
65 D x 43 mm H (2.6 x 1.7"); Probe: 4.5 D x 50 mm L
(0.18 x 2.0")
OM-MLT-LINK
PC INTERFACE FOR OM-MLT
SERIES DATA LOGGERS
PC Interface: data logger cradle
and 1.5 m (5') RS-232 cable with
DB9F termination Weight: 134 g (4.7 oz) Dimensions:
85 D x 29 mm H
(3.3 x 1.1")
OM-MLT-ACLINK
SERIAL INTERFACE ADAPTOR
FOR OM-MLT-AC
AUTOCLAVE DATA LOGGER
(used in conjunction with OM-MLT-LINK PC Interface
Weight: 124 g (4.4 oz) Dimensions:
85 D x 26 mm H
(3.3 x 1.0")
† All dollar amounts on this site are shown in US currency.
Note: Models OM-MLTXD16 includes one external temperature sensor. Model OM-MLTXXD16 includes two external temperature sensors. Ordering Example: (1) OM-MLT-KEY Starting magnet, $8.00