Economical Fiber Optic Infrared Temperature Transmitter with 4-20 mA Output |
| OS1553-A |
|
 |
|
|
Fiber Optic Infrared Temperature Transmitter
 |
| Click for larger image. |
|
|
$
2,175.00
OS1553-A-R1-01
|
| |
Reaches Difficult-to- Access Locations
|
| |
Fast, Precise 10 msec Time Constant
|
| |
Ideal for Small Targets with Spot Sizes Down to 0.004" Available
|
| |
Adjustable Emissivity for High Accuracy
|
| |
Spectral Filtering Allows Use in Dirty Atmospheres
|
| |
Works with Standard Process Meters and Controllers
|
View related products - Infrared Temperature Measurement Instruments
|
| |
|
|
|
 |
The OS1553-A is a self-contained fiber optic infrared transmitter designed to provide a linear 4 to 20mA output for monitoring, datalogging or controlling temperature.
The detector maybe remotely mounted up to 4' away from the fiber optic lens via the fiber optic cable, removing it from hazardous locations. Other problems that can be overcome by the OS1553-A fiber optic design include: taking measurements in high energy fields measuring very small objects or surfaces; and taking measurements in “dirty” atmospheres containing water vapor, dust or smoke.
Some important factors that will determine the fiber optic assembly needed for your infrared application are: the distance from the fiber optic lens to the surface being measured; the size of the object being measured; temperature range; ambient temperature around the fiber optic cable; the length of the cable; the method of heating the surface being measured; and the atmospheric conditions between the fiber optic lens and product.
SPECIFICATIONS
Accuracy: ±1.0% rdg or 1.7°C (±3°F), whichever is greater
Spectral Response: 0.95 to 1.09 microns
Repeatability: ±1°C/F or less
Time Constant: 10 msec (0 to 63.2%)
Output: 4 to 20 mA linear output
Power: 20 to 40 Vdc, 100 mA regulated, 0.2%
Power/Output Cable: 4 conductor shielded [4.5 m (15') standard]
Maximum Loop Resistance: Vsource-20/20 mA
Emissivity: 0.05 to 0.99, adj in 0.01 steps
Ambient Temperature: 0 to 50°C (32 to 122°F)
Dimensions: 69 Dia. x 203 mm L (2.7 x 8")
Weight: 1.6 kg (3.5 lb)
Fiber Optic Cable: 1.2 m (4') glass fiber material, heavy-duty stainless steel
|
| |
 |
|
 |
|
 |
† All dollar amounts on this site are shown in US currency.
Note: Comes complete with probe assembly, fiber cable, and operator’s manual. Consult OMEGA for information on custom OS1553-A fiber optic systems for any applicationb
Ordering Example: (1) OS1553-A-R1-02 Fiber optic sensor with 565 to 840°C (1050 to 1550°F) temperature range, .25" spot, $2175
|
|
 |
 |
 |
 |
 |
 |
 |
|
 |
 |
Part Number Builder |
 |
 |
 |
 |
 |
 |
|
|
|
Option Descriptions
|
|
(1)
|
Fiber Optic IR Transmitter
|
| |
Select OS1553-A-R1 for 565 to 840°C (1050 to 1550°F) OS1553-A-R2 for 816 to 1370°C (1500 to 2500°F)
|
(2)
|
Temperature Range
|
| |
Select 01 for 0.15" @ 6" - OS1553-A-R1 only 02 for 0.25" @ 8" - OS1553-A-R1 only 03 for 0.50" @ 16" - OS1553-A-R1 only 04 for 0.75" @ 24" - OS1553-A-R1 only 05 for 1.00" @ 30" - OS1553-A-R1 only 06 for 1.20" @ 36" - OS1553-A-R1 only 07 for .004" @ 1" - OS1553-A-R2 only 08 for .025" @ 4" - OS1553-A-R2 only 09 for .050" @ 10" - OS1553-A-R2 only 10 for .100" @ 18" - OS1553-A-R2 only 11 for .150" @ 24" - OS1553-A-R2 only 12 for .250" @ 36" - OS1553-A-R2 only
|
|
NOTE: All combinations may not be valid, check spec sheet for valid part numbers.
|
|
|
 |
|
|
 |
|  |
 |
 |
 |
 |
 |
 |
 |
 |
 |
Related Links -
Related Products
|
 |
 |
 |
 |
 |
 |
|
Related Products
|
 |
OS1592:
|
|
Fast Response Infrared Fiber Optic Thermometer/Transmitter with Local Display and Analog Output
|
 |
 |
|
 |
iR2P and iR2C:
|
|