High Temperature Insertion Type Electrode |
| PHE-5431-10 Series |
|
 |
|
|
High Temperature Insertion Type Electrode
 |
| Click for larger image. |
|
|
|
|
|
|
 |
The high temperature insertion type electrode is designed for periodic exposure to steam
sterilization or continuous
high temperature. This pH electrode is housed in a durable thermoplastic (PAS) body and is rated for
temperatures up to 135°C. The use of the double Porous PTFE liquid junctions with matched
viscosity electrolytes
provides a reference cell which permits extended periods of
pH measurements in the presence of sulfides or other
silver complexing agents. This electrode also incorporates pressure compensating
devices to protect the electrode
from extreme temperature induced pressure deviations.
Ten foot cable length and BNC connector are standard.
Application
• High Temperature Environment
• Continuous Processing Applications
• Harsh Conditions
• Steam Sterilization
Specifications
pH Range: 0 to 14 pH
Temperature Range:
-5 to +135°C (25 to 275°F)
+25°C to +110°C (76 to 230°F)
(High Temperature version)
Pressure Range: 500 psig @ 25°C
Accuracy: ±0.1% over full range
Impedance:
125 Mohm (standard version)
Reference Cell: Double Junction, KCl/AgCl, KNO3
Reference Junction: Porous PTFE
Wetted Materials: PTFE, Glass membrane, polymer outer body, EPR “O” Rings
Response Time: 95% of reading in 10 seconds
Drift: Less than 2 mV per week
|
| |
 |
|
|
MOST POPULAR MODELS HIGHLIGHTED! |
|
 |
|
 |
† All dollar amounts on this site are shown in US currency.
Ordering Example: (1) PHE-5431-10-T T-handle high temperature pH electrode, $185
|
|
 |
|
 |
|
 |
|
Back to top
|