The PHE-5432 is a steam-sterilizable combination pH electrode designed to
withstand high temperatures and pressures. The use of a porous PTFE liquid
double junction and specifically formulated low impedance glass membrane allows
it to function in the broadest spectrum of pH applications. This combination of
features permits extended periods of pH measurement in the presence of
poisoning ions or where membrane leaching would shorten the life of a
conventional electrode. Melting of the cable due to contact with steam lines and
motion generated noise are common problems with interconnecting cables on
steam sterilizable electrodes. Our proprietary TPE high temperature/ultra-low
noise cable is designed for optimal service, even in this demanding environment.
SPECIFICATIONS pH Range: 0 to 14 pH Temperature Range: -5 to 135°C (23 to 275°F) @ 25 psig Maximum Pressure: 500 psig @ 25°C (77°F) Accuracy: ±0.1% over full range Sodium Error: Less than 0.05 pH in 0.1 Molar Na+ ion @ 12.8 pH Reference Cell: Double Junction KNO3 and KCl/AgCl Reference Junction: Porous PTFE Zero Potential: 7.0 pH ±0.2 pH Wetted Materials: PTFE, glass membrane, glass outer body. Drift: Less than 2 mV per week
Model
PHE-5432-10
ORE-5432-10
Range
0 to 14 pH
/td>
±1999mV
Junction
PTFE Double
PTFE Double
Length
127mm/5.0''
130mm/5.1''
Pressure
500 psi @ 25°C
500 psi @ 25°C
Fittings Selection Guide
Model No.
Description
Model No.
Description
PHEH-51
1/2" MNPT, 316 Stainless Steel
PHEH-51F
1'' to 11/2" Sanitary Flange, 316SS
PHEH-51S
3/4" MNPT, 316 Stainless Steel
PHEH-52F
2'' Sanitary Flange, 316SS
PHEH-52
1" MNPT, 316 Stainless Steel
PHEH-51PG
25mm Port Gland 316 Stainless Steel
PHEH-54F
4'' Sanitary Flange, 316SS
† All dollar amounts on this site are shown in US currency.
Note: Comes complete with electrode care instruction sheet and 3 m (10') of cable.
For non-standard cable lengths, change the "-10" in the part number to the required length in feet, add $2 per foot to price (additional per foot cost applies to the entire length of cable.
For HF/Fluoride Resistant electrode, add suffix "-HF" to the model number, add $50 to price.
For 3000Ω Balco, add suffix "-R3K" to model number, add $50 to price.
For European PG style connector, add suffix "-PG" to the model number, add $50 to price.
NOTE: PHEH-51 (1/2 MNPT) installation fitting is required for first time installation.