OMEGA’s PX91 Amplified Output Series is a rugged, high temperature/ ultra high pressure transducer. It is hermetically sealed, has all stainless steel construction and is designed for the harshest environments encountered in very high pressure/ high temperature test chambers. The PX91 is available in sealed gage pressures from 5,000 psi to 200,000 psi. Electrical connection is either a sealed Bendix connector or lightning protected amplifier housing. Pressure connection is via a high pressure ferrule connection.
SPECIFICATIONS Excitation: 10 to 40 Vdc Output: 4 to 20 mA ±10% adj. Zero Balance: 4 mA +10% -2% adj Linearity: ±0.35% FSO <75K psi ±0.50% FSO ≥75K psi Hysteresis: ±0.10% FSO Repeatability: ±0.10% FSO Operational Temp Range: -46 to 121°C (-50 to 250°F) Compensated Temp Range: +17 to 71°C (+60 to 160°F) Thermal Effects: Span: 0.005% Rdg/°F Zero: 0.005% FSO/°F Proof Pressure: 5K to 50K psi: 120% of rated pressure 60K to 200K psi: 110% of rated pressure Burst Pressure: 5K to 75K psi: 150% of rated pressure 80K to 200K psi: 120% of rated pressure Wetted Parts: 5 to 75K: 17-4 PH Stainless Steel ≥80K: 455 Stainless Steel Pressure Port: 5K to 70K psi: 1/2-14 NPS (use 0.315 dia. 59° ferrule) >70K psi: 1-14 NS-2 (use 0.910 dia. 150° ferrule) Electrical Connection: Terminal Strip inside Lightning Protected Amplifier Housing
Examples of pricing for popular models is referenced in the "To Order" box. Use the "Part Number Builder" below to get pricing for all available models.
† All dollar amounts on this site are shown in US currency.
Note: Metric Ranges Available - Consult Engineering Ordering Example: (1) PX91N5-60KSI is a 60,000 psi sealed gage model with 4-20 mA output, a / 2 -14 NPS fitting and lightning protected amplifier housing, $860.00
Part Number Builder
Option Descriptions
(1)
Pressure Port
Select N for 1/2 -14 NPS use with 0.315in. dia 59° ferrule (70K psi or less) P for 1-14 NS-2 use with 0.910in. dia. 150° ferrule (greater than 70K psi)
(2)
Electrical Connection
Select 0 for 10 ft (3m) Cable 1 for PT1H-10-6P Connector or Equal (no charge) 5 for Lightening Protected Amplifier Housing (no charge)
(3)
Range (psi)
Select 5K for 0-5000 psi 10K for 0-10000 psi 15K for 0-15000 psi 20K for 0-20000 psi 30K for 0-30000 psi 35K for 0-35000 psi 50K for 0-50000 psi 60K for 0-60000 psi 80K for 0-80000 psi 100K for 0-100000 psi 120K for 0-120000 psi 150K for 0-150000 psi 200K for 0-200000 psi
(4)
Output
Select SV for mV/V S5T for 0-5 Vdc SI for 4-20 mA