Surface-Mount RTD Advanced-Design: Stick-On or Cement-On
Class A Accuracy |
| SA1-RTD Series |
|
 |
|
|
Surface-Mount RTD Advanced-Design Stick-On or Cement-On
 |
| Click for larger image. |
|
|
$
55.00
SA1-RTD
|
| |
100 Ω DIN Class A (±0.06 Ω or ±0.15°C at 0°C) Accuracy Standard
|
| |
Easy-Installation Silicone-Based, Self-Adhesive Backing Rated to 260°C (500°F)
|
| |
Back Can be Reapplied
|
| |
Temperature Range; -73C to 260°C Continuous, 290°C (554°F) Short Term Operation When Installed with OMEGABOND Air Set Cements
|
| |
Stripped 3-Wire Leads Standard (Connectors Optional), Color Code: Black/Black/Red
|
| |
Optional 4-Wire Configuration (Color Code: Black/Black/Red/Red)
|
| |
Stocked in 1 m (40") Lengths; Also Available in 2 and 3 m (80 and 120") and Custom Length Lead Wires
|
| |
Other Resistances/Accuracies Available on Request
|
| |
Stainless Steel Overbraid Option to reduce electrical noise and protect lead wire from abrasion
|
View related products - RTD (PT100) Probes, Elements and Assemblies
|
| |
|
|
|
 |
SPECIFICATIONS
Minimum/Maximum Temperature:
-73°C (-100°F) to 260°C (500°F) continuous,
290°C (554°F) short term (when cemented in place)
Sensing Element: 100 Ω at 0°C
(32°F), temp coefficient of 0.00385 Ω/Ω/°C (IEC60751)
Accuracy: ±0.06% at 0°C (DIN Class A)
Stability: Less than 0.2°C drift/year
Response Time: Less than 1sec. in water (63% response), less than 2sec. on a hot plate
Self-Heating Effect: 2.5 mW/°C
Lead Wire: 1 m (40"), 26 AWG,
nickel-plated copper, PFA-insulated 3-wire, Color Code: Black/Black/Red
The SA1-RTD surface-mount RTD
temperature sensor mounts on flat
or curved surfaces and provides
Class A accuracy (See IEC 751 Class A/B Standard Tolerances Table, below) for critical
temperature monitoring applications.
Based on a 2 x 2 x 0.8 mm thin-film
platinum RTD and supplied with
bare 3-wire leads as a standard
configuration (connector optional), it
can be customized for use in a wide
variety of applications. The sensor
can be easily applied with its selfadhesive
backing, or positioned
permanently with OMEGABOND® cements.
Use the SA1-RTD to monitor chip,
heat sink, and environmental
temperatures in electronic devices; check piping or ducting temperatures;
monitor motor and transformer core
heat; test insulation capabilities; and
check other applications in which
surface and/or gradient temperatures
need to be monitored or controlled
Also Available with Stainless Steel Overbraid
For applications where electrical noise is prevalent or where sensor
lead wire is being routed around sharp edges, the stainless steel
overbraid option (-SB) reduces noise and safeguards
wire leads. Add "-SB" to the end of the part number listed in
the table on the spec sheet, and add $6/m ($2/ft).
| IEC
751 Class A/B Standard Tolerances | Temp. (°C) | Tolerance
Class | | Class
A* | Class B | | ±°C | ±Ω | ±°C | ±Ω | | -200 | 0.55 | 0.24 | 1.3 | 0.56 | | -100 | 0.35 | 0.14 | 0.8 | 0.32 | | 0.0 | 0.15 | 0.06 | 0.3 | 0.12 | | 100 | 0.35 | 0.13 | 0.8 | 0.30 | | 200 | 0.55 | 0.20 | 1.3 | 0.48 | | 300 | 0.75 | 0.27 | 1.8 | 0.64 | | 400 | 0.95 | 0.33 | 2.3 | 0.79 | | 500 | 1.15 | 0.38 | 2.8 | 0.93 | | 600 | 1.35 | 0.43 | 3.3 | 1.06 | *SA1-RTDs
are IEC 751 Class A Devices.
|
| |
 |
|
 |
|
 |
† All dollar amounts on this site are shown in US currency.
Note: For special lengths of lead wire, add $2.25/ft. For an SA1-RTD with TA4F connector installed (SA1-RTD-4W-TA4F), $66; SA1-RTD-80-TA4F, $72; SA1-RTD-120-TA4F, $81. For more information, see the spec sheet or contact Sales.
(OPTIONS SHOWN IN THIS "NOTE" ABOVE ARE PRICED IN US$)
Ordering Example: (1) SA1-RTD-80-MTP Class A, surface-mount RTD sensor, 3-wire leads, 2 m (80") lead-wire length with a miniature 3-pin MTP connector Class A, surface-mount RTD sensor, 4 wire leads, 2 m (80") lead-wire length with a 4-pin, $70.00
|
|
 |
|
 |
|  |
 |
 |
 |
 |
 |
 |
 |
 |
 |
Related Links -
Related Products
|
 |
 |
 |
 |
 |
 |
|
Related Products
|
 |
TA3 and TA4:
|
|
Miniature Connectors for RTD's and Thermistors Series "T"
|
 |
 |
|
 |
MTP Series:
|
|
3-Prong Mini flat pin Connector for Thermocouple, RTD and 3-Wire Thermistor
|
 |
 |
|
|
|
|
|
|
 |
 |
|
|