Mini Temperature Transmitter With M12 Connectors |
| TX-M12-RTD Series |
|
 |
|
|
Mini Temperature Transmitter With M12 Connectors
 |
| Click for larger image. |
|
|
$
175.00
TX-M12-RTD-C-1
|
| |
M12 Connections Provide Easy Attachment to Your Sensor and Instrumentation
|
| |
Can be used with 2-Wire, 3-Wire or 4-Wire RTD Sensors (See Wiring Information Below)
|
| |
Programmable for use with Pt100, 500 or 1000 ohm RTD Sensors
|
| |
Small Size, 38mm Diameter Housing Weighs Just 100 grams (3 1/2 oz)
|
| |
4 to 20mA or 0 to 10VDC Output Models Available.
|
| |
Rangeable from -200 to 850°C (-328 to 1862°F)
|
View related products - RTD (PT100) Probes, Elements and Assemblies
|
| |
|
|
|
 |
Omega’s TX-M12-RTD Series transmitters offer
improved performance over conventional in-head
transmitters yet are a fraction of their size and weight.
Integral M12 connectors maintain IP67 protection and
connection integrity while allowing for a quick and
simple change of sensor. Two models are available
with either 4 to 20 mA or 0 to 10 Vdc output. Default
scaling for the full output is 0 to 100°C (32 to 212°F),
other ranges can be scaled at time of purchase for an
additional charge; or the optional USB programming
module (USB-CONFIG-UNIT) will allow the end user to
perform scaling.
The TX-M12-RTD transmitter is ideal for use with
Omega’s 4-wire PR-21, PR-22 and PRS-M12 Series RTD probes, or any
RTD probes with an M12 “A” coded pin style connector
which connects directly to theTX-M12-RTD sensor
input. Other 2-wire, 3-wire or 4-wire RTD probes, as well as slide wire
resistance devices, can be easily connected to the TX-M12-RTD transmitter
by using a field mountable M12-R-M-FM or M12-S-M-FM plug style
connector. Please see the wiring diagrams below for 2-wire, 3-wire and 4-wire
RTDs, as well as other slide wire resistance devices.
Don't forget to order an M12 extension cable to connect
your transmitter to your process instrumentation.
SPECIFICATIONS
Update Time: 200 mS
Response Time: 0.5 seconds
Warm-up Time: 1 minute to full accuracy
Sensor Connection: 2, 3 or 4-wire M12
Minimum Span: 25°C Thermal Drift
Pt100: 0.013 Ω/°C
Pt500/Pt1000: 0.063 Ω/°C
Accuracy: ±0.2°C + (0.05%) + output accuracy
Excitation Current: <200 uA
Lead Resistance Effect: 0.002°C/Ω
Maximum Lead Resistance: 20 Ω per leg
Operating Ambient: -40 to 85°C; 10 to 90% RH (non condensing)
Configuration Ambient: 10 to 30°C
Weight: Approx 100 g 4 to 20 mA Output Model
Output Type: 2-wire, 4 to 20 mA current loop
Output Range: 4.0 to 20.0 mA
Output Connection: M12 connector
Output Limits: 3.8 to 21.5 mA
Accuracy: mA output/2000 or 5 uA (whichever is the greater)
Loop Voltage Effect: 0.2 uA/V
Thermal Drift: 1 uA/°C
Maximum Output Load: [(Vsupply-10)/20] KΩ
Start Up Time 4 Seconds: I out < 4 mA during start up
Power Supply: 8 to 30 Vdc 0 to 10 Vdc Output Model
Output Type: 3-wire voltage
Output Range: Selectable within 0 to 10V
Current Drive: 2 mA max
Output Connection: M12 connector
Output Limits: -0.1 to 10.5V
Accuracy: Voltage output/1000 or 5 mV (whichever is the greater)
Thermal Drift: 1 mV/°C
Minimum Output: Load 5000 Ω
Power Supply: 12 to 30 Vdc
| TRANSMITTER SELECTION TABLE |
| Transmitter Model |
Transmitter Output |
Compatible Omega Sensors |
| TX-M12-RTD-V | 0 to 10VDC | PR-21…-2, PR-22, P-L-M12 |
| TX-M12-RTD-V-1 | 0 to 10VDC | PR-21…-1, PRS-M12, PR-25AP |
| TX-M12-RTD-C | 4 to 20mA | PR-21…-2, PR-22, P-L-M12 |
TX-M12-RTD-C-1 | 4 to 20mA | PR-21…-1, PRS-M12, PR-25AP |
| Sensor Input Wiring Diagrams |
 TX-M12-RTD-C and TX-M12-RTD-V |
|
 TX-M12-RTD-C-1 and TX-M12-RTD-V-1 |
|
| |
 |
|
 |
|
 |
† All dollar amounts on this site are shown in US currency.
Note: Comes complete with transmitter and operator’s manual (USB-CONFIG-UNIT interface cable sold separately).
Ordering Example: (1) TX-M12-RTD-C-1 RTD Transmitter with 4 to 20 mA Output For use with PR-21, PRS-M12 and PR-25AP Sensors, $175.00, plus (1) TX-M12-RTD-V-1 RTD Transmitter with 0 to 10VDC Output For use with PR-21, PRS-M12 and PR-25AP Sensors, $195.00, $175.00 + 195.00 = $370.00
|
|
 |
 |
 |
 |
 |
 |
 |
|
 |
 |
Part Number Builder |
 |
 |
 |
 |
 |
 |
|
|
|
Option Descriptions
|
|
(1)
|
Output Configuration
|
| |
Select C for 4 to 20mA Output V for 0 to 10VDC Output
|
(2)
|
Wiring Options
|
| |
Select No Dash for See Sensor Wiring Diagrams Above -1 for See Sensor Wiring Diagrams Above
|
|
NOTE: All combinations may not be valid, check spec sheet for valid part numbers.
|
|
|
 |
|
|
 |
|  |
 |
 |
 |
 |
 |
 |
 |
 |
 |
Related Links -
Related Products
|
 |
 |
 |
 |
 |
 |
|
Related Products
|
|
|
|
|
|
|
 |
M12CFM RTD Series:
|
|