The PXM509 intrinsically safe transducers are rugged all stainless steel transducers specifically designed to meet the requirements for use in hazardous locations for all classes and zones. Features include welded stainless steel construction, user accessible potentiometers, and a demountable electrical termination end. They are designed for rugged industrial applications, and have high shock and vibration ratings and excellent long term stability. A precision micromachined silicon sensor is at the core, and provides a very stable reading with exceptional high accuracy of 0.08% and a broad compensated range of -29 to 75°C (-20 to 167°F) on most ranges. Modular construction allows for fast delivery of most configurations and fittings including agency approved models, plus, customized models also can be delivered quickly to suit your exact needs. Models with ½ NPT conduit fitting (PX509C) do not have a demountable termination and trim pots are an extra cost option. This latest addition of the FM Approved Intrinsically Safe version for Hazardous Locations (available with 4-20 mA output) only further expands the range of applications of this versatile product family.
Recognized and respected across the globe, FM Approvals certification assures customers a product or service that has been objectively tested and conforms to the highest national and international standards. Please refer to FM Approvals for downloadable copies of FM Approvals Certificate of Compliance and Installation Control Drawings.
The PXM509-IS is also available in standard configurations, with ranges from 0-10 in H2O to 0-3500 psi ranges and standard threads. Please see PX509-IS for complete details, or contact our Pressure department for more information.
SPECIFICATIONS Compensated Temperature & Thermal Effects: Compensated Temperature 10 inH2O to 5 psi: -17 to 80°C 15 to 3500 psi: -29 to 80°C Zero Temperature Shift 10 inH2O to 5 psi: ±1.0% span 15 to 3500 psi: ±0.5% span Span Temperature Shift 10 inH2O to 5 psi: ±1.0% span 15 to 3500 psi: ±0.5% span Over Pressure10 inH2O: 10 times 1 psi: 6 times 2.5 to 500 psi: 4 times 750 to 3500: 3 times Containment 10 inH2O to 5 psi: To 1000 psi 15 to 750 psi: To 3000 psi 1000 psi: To 4000 psi 1500 psi: To 6000 psi 2500 psi: To 8000 psi 3500 psi: To 11,000 psi Absolute/Barometric Pressure Ranges Over Pressure 5 psia: 6 times span 15 to 1000 psia: 4 times span 1500 to 5000 psia: 7250 psia maximum Containment All Ranges: To 10,000 psi Gauge/Vacuum/Compound Pressure Ranges Over Pressure 10 inH2O: 10 times span 1 psi: 6 times span 2.5 to 1000 psi: 4 times span 1500 to 5000 psi: 7250 psi maximum Minimum Resistance Between Transducer Body and Any Wire: 100MΩ @ 50 Vdc Accuracy (Combined Linearity, Hysteresis and Repeatability): ±0.08% BSL maximum (other accuracies available) For conduit (C and CW) zero and span set ±0.5% full scale typical ±1% maximum or set ±1% full scale typical ±2% maximum ranges 2.5 psi and below Pressure Cycles: 1 million typical Long Term Stability (1-Year): ±0.1% full scale typical Supply Voltage: 10 to 28 Vdc Output: 4 to 20 mA Operating Temperature: -40 to 80°C (-40 to 176°F) Wetted Material: 316L stainless steel Bandwidth: DC to 1kHz (typical) Sensor: Oil filled silicon Shock: 50 g, 11 msec half sine shock Vibration: ±20 g Response Time: <1 msec RoHS Compliant Weight: 315 g (11 oz) typical depending upon configuration