text.skipToContent text.skipToNavigation

Precision RTD Handheld Data Logger Thermometer

$372.58

Item# HH376

 This item is available for quick shipping
Volume discounts available
Volume Discount Schedule
Quantity Price each
1-4 $372.58
5-9 $353.95
10-24 $346.50
25-49 $331.60
50-99 $324.14
100-199 $316.69
200+ $312.32

In stock

Lead Time ( If not in Stock): 2 weeks

Precision RTD Handheld Data Logger Thermometer HH376 HH376
Precision RTD Handheld Data Logger Thermometer is rated 5.0 out of 5 by 3.
  • y_2020, m_8, d_13, h_15
  • bvseo_bulk, prod_bvrr, vn_bulk_3.0.10
  • cp_1, bvpage1
  • co_hasreviews, tv_0, tr_3
  • loc_en_US, sid_HH376, prod, sort_[SortEntry(order=HELPFULNESS, direction=DESCENDING), SortEntry(order=POSITIVE_FEEDBACK, direction=DESCENDING), SortEntry(order=SUBMISSION_TIME, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • LOCAL, getAggregateRating, 2ms
  • REVIEWS, PRODUCT
  • High Accuracy with 0.01C/F Resolution
  • Large Backlit Dual LCD Display
  • MAX/MIN/HOLD Function
  • 16,000 Record Data Logger
  • USB Cable Included
  • Windows Software
  • Auto Power Off
  • User Offset Adjustable
  • Low Battery Indicator
  • Tilt Stand
  • Precision 4 Wire 100 Ω Platinum RTD Probe Included
5 Year Warranty

Edit Options

View all models

Not all combinations are valid. Options compatible with previous selections will be in bold.

*Highlighted options are not compatible. Please select a different combination.

Category
Input Type
Minimum Temperature Measurement
Maximum Temperature Measurement
Input Connection Type
Display Type
Accuracy Value
Resolution
Supply Power
Number of channels
Probe Size Cable Length
Storage
Communications Protocol
Features


Product Specs

  • Battery life Approx. 100 hrs
  • Battery Size 9V
  • Maximum Ambient Humidity 80% RH
  • Maximum Storage Temperature 60°C (140°F)
  • Measurement Units °F/°C
  • Minimum Ambient Humidity 0% RH
  • Minimum Storage Temperature -10°C (14°F)
  • Number of Batteries Included 1
  • Power Source Battery
  • Maximum Ambient Temperature 50°C (122°F)
  • Sample Rate 2 times per second

The HH376 is a high accuracy RTD thermometer plus a powerful data logger. The unit can store up to 16,000 records at programmed intervals. Windows® based software and USB interface cable are included to allow real time or recorded data to be read by a PC through the USB or RS232 port (RS232 cable is optional). HH376 is supplied with a precision 4-wire RTD probe with automatic locked push-pull connector for easy connection.

SPECIFICATIONS
Instrument
Numerical Display:
5-digit LCD
Measurement Range: -100 to 400°C (-148 to 752°F)
Resolution: 0.01°C (0.02°F)
Sensor Types: Platinum resistance temperature sensor for pt-100 4 wires, ALPHA = 0.00385

Environmental
Operating Temperature and Humidity:
0 to 50°C (32 to 122°F); 0 to 80% RH
Storage Temperature: -10 to 60°C (14 to 140°F); 0 to 80% RH; altitude up to 2000 meters
Accuracy: At (23 ±5°C)
Range Accuracy: -100 to 400°C ±(0.05% reading + 0.1°C); -148 to 752°F ±(0.05% reading + 0.2°F)
Temperature Coefficient: For ambient temperatures from 0 to 18°C and 28 to 50°C, for each °C ambient below 18°C or above 28°C add the following tolerance into the accuracy spec. 0.005% of reading + 0.01°C (0.005% of reading + 0.02°F)
Sample Rate: 2 times per second
Dimensions: 185 L x 65 W x 36 mm D (meter only)
Weight: 360 g approx.
Power Requirement: 9V battery (included)
AC Adaptor: 9 Vdc (7 to 10V max)/20 mA
min (plug diameter: 3.5 x 1.35 mm)
Battery Life: Approx. 100 hrs

Probe
Sensor Type:
Pt100 (4 wire)
Accuracy: Class A13, temperature probe
Measurement Range: -100 to 400°C (-148 to 752°F)
Sheath Diameter: 5.0 mm (0.2')
Sheath Length: 230 mm (9.05')
Cable Length: 1500 mm (59.0')

Rated 5 out of 5 by from Very accurate instrument After using numerous instruments to measure pipeline temperatures, I have finally found the device that accurately does.
Date published: 2017-11-28
Rated 5 out of 5 by from High accuracy and inexpensive Great little handheld with 2 digits of precision after the decimal. Data logging for charting responses, easy connect to computer for download of data. We use it to monitor differential temps near 100C and it’s very repeatable. Also nice is the 120V adapter so no worry about battery life. Highly recommend.
Date published: 2018-06-09
Rated 5 out of 5 by from Price quality Price Quality, Practical dan easy to use. I will give good use.
Date published: 2018-01-08
  • y_2020, m_8, d_13, h_15
  • bvseo_bulk, prod_bvrr, vn_bulk_3.0.10
  • cp_1, bvpage1
  • co_hasreviews, tv_0, tr_3
  • loc_en_US, sid_HH376, prod, sort_[SortEntry(order=HELPFULNESS, direction=DESCENDING), SortEntry(order=POSITIVE_FEEDBACK, direction=DESCENDING), SortEntry(order=SUBMISSION_TIME, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • LOCAL, getReviews, 0ms
  • REVIEWS, PRODUCT

Can you explain the accuracy spec? It reads 23+/-5°C, with no spec, then a range spec for -100 to +400°C.

Asked by: RPMS
The 23 +/-5 C is the ambient temperature the meter needs to be in to meet the accuracy specification over the measurement range.
Answered by: ANB ENG
Date published: 2020-07-13

At (23 ±5°C) : Range Accuracy: -100 to 400°C ±(0.05% reading + 0.1°C); -148 to 752°F ±(0.05% reading + 0.2°F) Is accuracy unit and probe, or unit only?

Asked by: DavNH
Hi DavNH, Thank you for the inquiry. Accuracy is for the meter only. Probe is accuracy class A ( 0.15 C +/- 0.2 % of reading) Best regards OMEGA
Answered by: NorwalkAE
Date published: 2020-05-27

Sampling rate is 2 per seconds, what is the logging interval for this device?

Asked by: Chase
Thank you for your inquiry. The logging interval minimum 1 second.
Answered by: Engineer CB
Date published: 2020-05-06

NIST traceable option?

Asked by: Bumbleguy
Thank you for your inquiry. Unit is supplied with a NIST certificate (No Points). For an additional cost you can order with a NIST certificate with points.
Answered by: Application Engineer JB
Date published: 2020-04-06
  • y_2020, m_8, d_13, h_15CST
  • bvseo_bulk, prod_bvqa, vn_bulk_3.0.10
  • cp_1, bvpage1
  • co_hasquestionsanswers, tq_4
  • loc_en_US, sid_HH376, prod, sort_[SortEntry(order=LAST_APPROVED_ANSWER_SUBMISSION_TIME, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • LOCAL, getContent, 1ms
  • QUESTIONS, PRODUCT
Accessories
Precision 4 wire RTD probe (spare)
$88.78
1 In Stock
Power adaptor (115 Vac)
$19.75
7 In Stock
RS232 interface cable
$19.75
23 In Stock
Spare USB interface cable
$19.75
29 In Stock

en-US
en
Save Configuration