Portable Oxygen Monitor and Data Logger is rated
out of
5 by
1.
- Large LCD displays O2 level, Pressure, Temperature and Relative Humidity
- Calendar (Y/M/D) and Time
- Built in Fan for Fast Sampling
- Stable Electrochemical Sensor for O2 Detection
- Altitude Compensation
Edit Options
View all modelsNot all combinations are valid. Options compatible with previous selections will be in bold.
*Highlighted options are not compatible. Please select a different combination.
Measurement Type
Range
Accuracy
Supply Power
Features
Product Specs
The HHAQ-104 is designed to measure O2 concentration, pressure, air temperature and relative humidity with visible and audible alarms. This O2 monitor is an ideal instrument for indoor air quality (IAQ) diagnosis and HVAC system performance verification. The HHAQ-104 features a remote probe and detachable soft in/out tube to reach air-ducts or where access with the probe is not possible.
SPECIFICATIONS
O2
Measurement Range: 5.0 to 30.0%
Accuracy: ±1%
Resolution: 0.1%
Temperature
Range: -10 to 60°C (14 to 140°F)
Accuracy: ±0.6°C (0.9°F)
Resolution: 1°C (77°F)
Relative Humidity
Range: 0.1 to 99% RH
Accuracy: ±3% RH [at 25°C (77°F), from 10 to 90% RH], ±5% RH above and below
Pressure Range: 30 to 110 kPa
Memory Capacity: 32,000 readings
Storage Temperature: -40 to 70°C (-40 to 158°F)
Interface: USB
Response Time: @ 2 minute to 100%
Dimensions:
Meter: 160 H x 60 W x 40 D mm (6.30 x 2.35 x 1.5")
Probe: 142 H x 32 W x 34 D mm (5.6 x 1.25 x 1.33")
Weight: 550 g (19.4 oz)
SPECIFICATIONS
O2
Measurement Range: 5.0 to 30.0%
Accuracy: ±1%
Resolution: 0.1%
Temperature
Range: -10 to 60°C (14 to 140°F)
Accuracy: ±0.6°C (0.9°F)
Resolution: 1°C (77°F)
Relative Humidity
Range: 0.1 to 99% RH
Accuracy: ±3% RH [at 25°C (77°F), from 10 to 90% RH], ±5% RH above and below
Pressure Range: 30 to 110 kPa
Memory Capacity: 32,000 readings
Storage Temperature: -40 to 70°C (-40 to 158°F)
Interface: USB
Response Time: @ 2 minute to 100%
Dimensions:
Meter: 160 H x 60 W x 40 D mm (6.30 x 2.35 x 1.5")
Probe: 142 H x 32 W x 34 D mm (5.6 x 1.25 x 1.33")
Weight: 550 g (19.4 oz)
Rated 4 out of
5
by
Mark2 from
Easy to use
Found that this logger was very easy to use. The instructions were sparse, but adequate to run the logger. I was unable to update the clock on the logger with the correct date and time. It would update when running 'real time' output direct to computer, but would not log correct time when isolated from the computer.
Date published: 2018-01-22
What is the rated lifespan of the sensor? Ie, how long is the sensor good for until it needs to be replaced?
Thank you for your inquiry. Replace the sensor on an annual basis.
Date published: 2020-12-17
is these equipment made in Canada? Also if it need calibration are capable of calibrating and providing certificate of calibrations and how long does it take? I am in the Greater Toronto Area
In regard to your inquiry, it is not made in Canada and unfortunately, there is no NIST option available. Omega can supply a Certificate of Conformance, or CAL-1, which is free of charge.
Date published: 2019-11-21
I want to hook this unit up to our medical air system which runs at 50psi. I want to look for variation in oxygen levels in our medical air. Strictly for troubleshooting purposes and not certification. Would this unit be suitable?
No. This meter has no pressure rating. 50 PSI will hurt the meter.
Date published: 2020-02-20
Is this instrument intrinsically safe?
Hi, thanks for your question to Omega. The HHAQ-104 series is not rated intrinsically safe. Please connect with us if you have any further questions.
Date published: 2019-06-26