The CDH-93 resistivity and conductivity meter uses advanced automatic frequency conversion and booster technology to keep the electrode constant the same in different solution strengths. It can meet measuring accuracy in the whole range of this electrode by using only the one-point calibration function. The result is the high performance and the lowest cost.
Specifications Range Conductivity: 5 to 50.0µ S/cm, 50 to 500µ S/cm, 0.5 to 5.00 mS/cm, 5.0 to 50.0 mS/cm Resistivity: 2 to 20 M?-cm TDS: 0 to 35.0 g/L Salinity: 0 to 25.0 ppt Temperature: 0 to 50°C (32 to 122°F) Accuracy: Conductivity/Resistivity: ±1.0% full scale temperature; ±0.5°C ATC Range: 0 to 50°C (32 to 122°F) Calibration: One-point calibration (1413µ S/cm) Power: Two (2) CR2032 3V lithium ion batteries (included) IP Standard: Waterproof and dustproof meets NEMA 6 (IP57) standard Auto Power Off: After 10 minutes (override available) Dimensions: Meter: 148 L × 29 W × 14 mm D (5.82 x 1.14 x 0.55") Carrying Case: 168 L × 96 W × 28 mm D (6.61 x 3.78 x 1.10") Weight with Case: 178 g (6.28 oz)
Rated 4 out of
5 by
samm from
Good Tool. Poor instructions. So-so support.This will be a good tool for the intended purposes. It is easy to use and appears accurate and durable, but the documentation that comes with it is very meager. I called with a few questions. One was answered other wasn't. Didn't receive a return call or a response to an online support question submitted as a follow up.
Date published: 2017-05-02
Rated 1 out of
5 by
analytical chemist from
Nonsensical rangesThe advertised range for this probe is 5 uS/cm - 50 mS/cm, which converts to 20 - 20000 Ohm*cm (0.020 - 200 KOhm*cm), yet they claim it has a resistivity range of 0-20 MOhm*cm. This would require conductivity readings to go as low as 0.05 uS/cm. In practice, I can't seem to get a single resistivity reading higher than 3 KOhm*cm. I am working with deionized water that should be around 18 MOhm*cm. This is completely unacceptable equipment for the purposes I needed and if you also need to confirm deionized water purity, do not waste your money on this.
Date published: 2023-01-24
Rated 5 out of
5 by
Mack Harris from
Wire EDMI use this tester to verify my water conductivity for wire EDM it easy to use and fast
Hi. Our spec for resistivity is 2 M Ohm or greater. My question is, Does the meter just cut off and show no reading below 2MOhm? If it does read anything below 2M, does accuracy stay at 1% of full scale?
Asked by: Ed46
Than you your inquiry for this meter. Unfortunately, it will not be able read below 2M Ohm.
Answered by: JC ENGR
Date published: 2020-11-05
Do you sell a meter that can read 500 ohms to 20 Megohms?
Asked by: Satch
Thanks for you question.
Omega doesn't have a product that goes to the ranges you have specified.
please contact the flow department to see if they would have other options?
Answered by: application engineer je
Date published: 2020-02-27
Conductivity: 5 to 50.0µ S/cm, 50 to 500µ S/cm, 0.5 to 5.00 mS/cm, 5.0 to 50.0 mS/cm. The resolution is ±1% of full scale. Does that mean the accuracy is ±0.01x50mS/cm which is ± 0.5mS/cm?
Asked by: Maywu
Thank you for your inquiry about the handheld meter. Yes, this is correct for the accuracy.
Answered by: JC ENGR
Date published: 2020-09-09
Is your unit sufficient for testing DI water
Asked by: QM Mgr
Thank you for your inquiry. It would be advised to email our Flow, Level, and Environmental Group at flow@omega.com for that recommendation.
Answered by: JC ENGR
Date published: 2020-12-09
The range is 2 megaohms to 20 megaohms correct?
Do these come with calibration certificates?
How long are the calibrations good for and can we send them to you guys to calibrate periodically before calibrations expire?
Do you guys sell a standards?
Asked by: Anonymous
Thank you for your inquiry. Yes the resistivity range is 2 to 20 MΩ-cm. These electrodes are calibrated prior to shipping but calibration certificates are not provided. We can provide a Certificate of Conformance though. Field calibrations should be done at least monthly depending on frequency of use. We do sell calibration buffer solutions for this meter. The part number for the solution is CDSA-1413. For additional inquiries please contact us at flow@omega.com and our pH/Flow engineers will be able to help you.