Vortex Shedding Flow Meter and Temperature Transmitter
1/2" NPT Female Flow Only Transmitter
Volume discounts available
Volume Discount Schedule
Quantity | Price each |
---|---|
1-4 | $756.19 |
5-9 | $718.38 |
10-24 | $703.26 |
25-49 | $673.01 |
50-99 | $657.89 |
100+ | $642.76 |
Added to Your Shopping Cart
Configuration complete.
Please add to cart to keep it or exit the configuration
Please add to cart to keep it or exit the configuration
Failed to Save Configurations, Click on Configure Button Start Again
- No Moving Parts to Jam
- 1/4 FNPT to 4" Line Sizes
- Bright LED Display
- Individual 4 to 20 mA Outputs for Flow and Temperature
- SPST SSR Alarm(s)
- Learn more about the vortex flow meter principle.
FV102 Model Options
View all modelsEdit these specs to order a different model. Not all combinations are valid. Options compatible with previous selections will be in bold.
*Highlighted options are not compatible. Please select a different combination.
Flow Range
Process Connection Type
Process Connection Size
Wetted Materials
Max Operating Pressure
Output Signal
Features
Category
Accuracy
Supply Power
Max Process Temperature
Display
The FV101 Series comprises in-line flow meters that use the vortex shedding measuring principle. The fluid strikes a bluff body, imparting alternating vortices downstream of the bluff. This creates a pressure on a sensor body containing a piezoelectric crystal. The frequency of the sensor is proportional to the velocity of the fluid and is amplified and converted to a 4 to 20 mA output signal. Vortex technology yields a meter with no moving parts to hang up or wear. The meter has a bright 7.62 mm (0.3") high LED display of flow in either liters or gallons. The FV101 Series can be used with non-viscous, clean, or dirty waterlike liquids that are compatible with brass, PVDF, and FKM. Measured fluids should not contain long fibers or a high level of abrasive solids. These meters are ideal for cooling loops using water or 50% glycols, and for water-soluble machine coolant (up to 10%). These applications are found in most process industries, including rubber, steel, fabrication, manufacturing, refining, paper, chemical, food, petrochemical, and power. Do not use on flammable liquids or on gases such as air.
SPECIFICATIONS
Flow
Accuracy: ±2% FS
Repeatability: ±0.25% FS
Alarm Deadband: 5% FS
Response Time: User selectable from 0.9 to 7.5 seconds
Max Flow: Over range occasionally up to 125% of capacity
Temperature
Accuracy: ±1% FS
Response Time: 450mS time constant
Alarm Deadband: 2%
Display: 3-digit, 7.62 mm (0.3") high numeric LED that flashes below setpoint
Maximum Operating Pressure:
2" and Smaller: 20 bar (300 psig)
3 and 4": 13 bar (200 psig)
Operating Temperature: 2 to 66°C (35 to 150°F)
Enclosure Rating: IP65; Type 1, 3, 4, 12, and 13
Analog Output: 4 to 20 mA (600 ? @ 24 Vdc)
Electrical Connection:
FV100 Units (Flow Only): 5-pin micro male connector
FV100-T Units (Flow/Temperature): 8-pin male micro connector
Alarm Output: SPST SSR, NO or NC operation, selectable; dual alarms for flow/temperature units
FV101 and 102:250 mA @ 30 Vdc up to 66°C (150°F)
FV103 thru 108:125 mA @ 30 Vdc up to 66°C (150°F); reduced current rating at higher temperatures
Power: 10 to 30 Vdc @ 80 mA
Wetted Parts: Brass, PVDF and FKM seals; 316 SS bodies available for special order on 1/4 to 1/2" sizes, contact Flow Engineering for details
Installation: 10 pipe diameters upstream and 5 downstream recommended
SPECIFICATIONS
Flow
Accuracy: ±2% FS
Repeatability: ±0.25% FS
Alarm Deadband: 5% FS
Response Time: User selectable from 0.9 to 7.5 seconds
Max Flow: Over range occasionally up to 125% of capacity
Temperature
Accuracy: ±1% FS
Response Time: 450mS time constant
Alarm Deadband: 2%
Display: 3-digit, 7.62 mm (0.3") high numeric LED that flashes below setpoint
Maximum Operating Pressure:
2" and Smaller: 20 bar (300 psig)
3 and 4": 13 bar (200 psig)
Operating Temperature: 2 to 66°C (35 to 150°F)
Enclosure Rating: IP65; Type 1, 3, 4, 12, and 13
Analog Output: 4 to 20 mA (600 ? @ 24 Vdc)
Electrical Connection:
FV100 Units (Flow Only): 5-pin micro male connector
FV100-T Units (Flow/Temperature): 8-pin male micro connector
Alarm Output: SPST SSR, NO or NC operation, selectable; dual alarms for flow/temperature units
FV101 and 102:250 mA @ 30 Vdc up to 66°C (150°F)
FV103 thru 108:125 mA @ 30 Vdc up to 66°C (150°F); reduced current rating at higher temperatures
Power: 10 to 30 Vdc @ 80 mA
Wetted Parts: Brass, PVDF and FKM seals; 316 SS bodies available for special order on 1/4 to 1/2" sizes, contact Flow Engineering for details
Installation: 10 pipe diameters upstream and 5 downstream recommended
Flowmeter Model No. | Flowmeter/ Temperature Transmitter Model No. | Fittings | Flow Range (gpm) | Max Pressure Drop psid | Weight kg (lb) | |
Min | Max | |||||
FV101 | — | ¼ FNPT | 0.4 | 4 | 9.2 | 1.3 (2.8) |
FV102 | — | ½ FNPT | 1.2 | 12 | 3.6 | 1.2 (2.6) |
FV103 | FV103-T | ¾ FNPT | 2.5 | 25 | 3.6 | 2.9 (6.3) |
FV104 | FV104-T | 1 FNPT | 5 | 50 | 9.5 | 2.7 (5.9) |
FV105 | FV105-T | 1½ FNPT | 10 | 100 | 4.8 | 5.9 (13) |
FV106 | FV106-T | 2 FNPT | 20 | 200 | 4.9 | 4.8 (10.5) |
FV107 | FV107-T | 3" ANSI RF | 30 | 300 | 1.3 | 19 (42) |
FV108 | FV108-T | 4" ANSI RF | 60 | 600 | 1.3 | 25 (55) |



Accessories